Environmental Test Chamber CTC256 With Humidity And Temperature Control, 230V - 256L Capacity

Environmental Test Chamber CTC256 With Humidity And Temperature Control, 230V - 256L Capacity

(No reviews yet) (No reviews yet)
SKU:
MMT-CTC256
Weight:
743.00 LBS
Shipping:
Calculated at Checkout
$31,104.00
Share:

The Memmert CTC256 environmental testing chamber is quiet, space-saving, environmentally friendly, and extremely economical. 

Offering a total of 256L of testing space and controls (TFT display) that are so easy to use, you will be wondering how you ever managed without a Memmert!

Maintain incredible tight controls over internal humidity with the built in self-priming water pump and built-in hot steam generator.

De-humidification is done by means of a cold trap that uses Peltier technology for precision.

Electronic fuzzy-supported PID process controller with permanent performance adjustments and time-saving self-diagnosis system to quickly locate errors

Program sequence control for up to 40 ramp segments

Calibration certificates for temperatures at –20 °C and +160 °C and for humidity at 30 °C and 60 % rh

Feed-through 80 mm on right with silicon sponge plug

USB interface (optionally Ethernet) for computer-based temperature control program and to read out the controller’s internal log memory

The programming interface can be set in English, German, Spanish, French, Czech and Hungarian to accommodate a wide range of lab personnel. 

Automatically restores to last program in event of power failure

The AtmosControl software allow for remote programming that can be done via mouse/PC or tablet, and data can be transferred using USB stick and Ethernet/LAN. 

Heated inner glass panes with 2-point locking mechanism

Integrated over and undertemperature protection "ASF", automatically following the set-point value at a preset tolerance range, visual and audio alarm in case of over or undertemperature.

Important!

This test chamber is not explosion-proof (does not comply with the workplace health & safety regulation VBG 24 of the Administrative Employers‘ Liability Insurance Association). This testing chamber may only be loaded with materials and substances which cannot produce any toxic or explosive vapors at temperature ranges up to 190 ºC, and which themselves cannot explode, burst or ignite.

This  test chamber may not be used for drying, vaporizing and branding paints or similar materials, the solvents of which could form an explosive mixture when combined with air. If there is any doubt as to the composition of materials, they must not be loaded into the test chamber.

Potentially explosive gas-air mixtures must not be produced, either in the interior of the chamber or in the direct vicinity of the appliance.

Comes with one stainless-steel, electropolished shelf

Installed on locking casters

Temperature Range:
Without humidity: from -42°C to 190°C, With humidity: from 10°C to 95°C
Temperature Accuracy:
0.1°C
Humidity Range:
10 - 90 % rh
Humidity Accuracy:
0.5 %
Calibration:
Three freely selectable temperature values, 2-point calibration for humidity: 20 and 90 % rh
Interface:
Ethernet LAN, USB
Interior Dimensions (W x H x D):
25.2” x 26.4” x 23.5
Exterior Dimensions (W x H x D):
35.4” x 68.1” x 43.3”
Maximum Number of Shelves:
14
Maximum Loading Weight Per Shelf:
30Kg